01
Because the device replaces the ventricles rather than sitting inside them, the pump housing is larger — allowing radial blood gaps of 300 μm and axial gaps of up to 10,000 μm, roughly 10–20× wider than typical rotary pumps. Wider gaps are designed to lower shear stress on formed blood elements, with the goal of reducing hemolysis. Preliminary data suggest the TAH may preserve high molecular weight von Willebrand Factor multimers, which are degraded by high-shear pumps and drive acquired von Willebrand syndrome and GI bleeding. In testing, the device operates at full pulsatility across both circulations simultaneously.
02
The controller cyclically modulates rotor speed (900 RPM) to produce pulsatile outflow with pulse pressures up to 40 mmHg, palpable in carotid, femoral, and radial arteries. Pulse oximetry reads SpO₂ normally. Based on our current clinical trial protocol, no specialized monitoring equipment is required. Outpatient hemodynamic management follows a protocol consistent with native heart physiology.
03
Cyclic rotor acceleration and deceleration are designed to wash blood-contacting surfaces, an approach being studied for its potential to reduce thrombus formation at the pump interface. The right-side impeller features a paddle-like blade geometry designed to manage potential embolic material originating from the deep veins. Because both native ventricles are excised at implant, the dilated and akinetic myocardium — the primary structural substrate for intracardiac thrombus and embolic stroke — is physically removed from the circulation. The clinical implications of this anatomical design feature are under active evaluation in the EFS.
04
The rotor is suspended entirely by electromagnetic force. No mechanical bearings, no contact, no wear. Lab studies demonstrate the bearing system withstands forces of more than 4–6 G, similar to the accelerations of everyday activities. The titanium housing is approximately the same size as a normal human heart and 1.5× its weight, but smaller and lighter than the failing heart it replaces. Its size is designed to accommodate patients with a BSA of 1.4 m² or more, including most women and many adolescents, something not always offered by previous total artificial hearts.
05
The device is being designed to deliver cardiac output sufficient for rest (4–6 L/min) through vigorous physical activity, within the FDA-approved range of 3–12 L/min. Intrinsic preload and afterload sensitivity enables automatic output increases without clinician intervention. In the initial EFS cohort, most patients were ambulatory within the first week post-implant. Both ventricles are excised at implant, leaving no residual myocardium. The implications for medical therapy are being evaluated in the EFS. The device does not require immunosuppression.
06
The magnetic bearing system continuously senses rotor position, balancing left and right pump outputs in real time. When venous return increases during exercise or postural change, cardiac output rises automatically (4-5L/min) without clinician intervention. This intrinsic preload and afterload sensitivity, which responds to the body’s native vessel changes during daily activities, is designed to prevent underfilling and suction events associated with syncope in ventricular assist. The hemodynamic response to physiologic demand, as a healthy heart does, is a core design principle of the TAH architecture.
The TAH by BiVACOR is investigational and has not been approved or cleared by the FDA for commercial use.
07
Current protocol: warfarin, TEG-guided, specific protocol. Not ACT-based and not transferable from LVAD protocols. Given the device's hemocompatibility profile — low shear, wide blood gaps, comparable vWF multimer performance to reference pumps in vitro — alternative anticoagulation strategies are being evaluated for future clinical phases, including direct oral anticoagulants and potentially reduced anticoagulation intensity. Any protocol changes will be pursued through formal regulatory pathways with appropriate clinical evidence.